In 20 children and young adults with lymphoma, SubtlePET™ enabled correct treatment response assessment at half the [18F]FDG dose.
Children and young adults with lymphoma get repeat PET scans to track how their tumors respond to chemotherapy, and each scan adds radiation exposure. This study, published in Radiology: Artificial Intelligence, asked whether deep learning could support a lower [18F]FDG dose without changing the response assessment.
Researchers at Stanford studied 20 patients, average age 16, who had whole-body PET/MRI before and after induction chemotherapy. From each full-dose scan, they simulated lower doses down to 6.25%, then enhanced them with SubtlePET™. None of these scans had been used to train the software. For each follow-up scan, the team scored response by comparing tumor uptake with the liver and classified patients as adequate or inadequate responders.
Lower doses on their own caused a real problem. Noise pushed tumor SUVmax up, which made good responders look like poor ones. At a simulated 75% dose, 6 of 14 adequate responders were misclassified. At 50%, it was 13 of 14, and below that, all 14.
With deep learning enhancement, the 75% and 50% scans classified every patient correctly, matching the full-dose scans with 100% sensitivity and specificity.
The cohort was small and the lower doses were simulated, not injected. Still, the result shows why denoising matters here. The risk of a low-dose scan isn’t only a noisier image. It’s a misclassified response.
The authors conclude that deep learning enhancement may allow a 50% [18F]FDG dose reduction with correct response assessment in this population.