In 195 patients, AI denoising brought half-duration [18F]FDG PET to full-duration image-quality scores, with 97.7% lesion concordance.
Demand for PET keeps climbing, and long acquisitions mean fewer slots and more time on the table for patients. This prospective study, published in the European Journal of Nuclear Medicine and Molecular Imaging, asked whether AI denoising could make a half-length [18F]FDG scan perform like a full-length one on digital PET/CT.
Researchers at the Baclesse Cancer Center in Caen, France, enrolled 195 patients and acquired body PET in list mode on a Philips Vereos at 90 seconds per bed position. From the same data, they reconstructed a half-duration scan at 45 seconds per bed position, then processed it with SubtlePET™. Readers compared three versions: the original, the half-duration scan alone, and the half-duration scan with AI denoising.
Without denoising, the half-duration scan fell short. Image quality was scored poor in 8% of cases and moderate in 68%, and lesion concordance with the original was 88.7%.
With denoising, image quality matched the original: good in 92% of cases and moderate in 8% for both. Readers found 836 of 856 lesions on both scans (97.7%). Seven lesions appeared only on the original and 13 only on the denoised scan. Processing took 90 seconds on average.
Quantitative measures stayed closely aligned, with intraclass correlations of 0.97 or higher for lesion SUL and metabolic volume. Some values did shift. Lesion SULmax was lower on the denoised scan and metabolic volume was higher, while EARL1 SULpeak showed no significant difference.
The authors conclude that AI denoising allows [18F]FDG PET acquisition time to be halved on digital PET/CT. They note that readers compared images side by side, which may have helped detection, and call for multicenter studies with other radiotracers.