Clinical and phantom validation of a deep learning based denoising algorithm for F-18-FDG PET images from lower detection counting in comparison with the standard acquisition

Across 3 PET scanners and 110 patients, SubtlePET™ matched full-time image quality on half-count [18F]FDG PET at the European standard dose.

May 11, 2022

SubtlePET™ was cleared in the US for PET images acquired at 25% of standard count statistics, based on US injection practices. This study, published in EJNMMI Physics, tested where the algorithm’s limits sit when sites follow the European guideline of 3 MBq/kg of [18F]FDG.

Researchers in Rouen, France, started with phantoms on 3 GE scanners: the Discovery MI, Discovery 710, and Discovery IQ. They imaged spheres at low and high contrast, plus a set of micro-spheres, then moved to 110 patients with a range of pathologies. All data was acquired in list mode, so the team could reconstruct each scan at full counts, half counts, and one-third counts from the same acquisition. The reduced scans were then denoised with SubtlePET™. Two senior nuclear medicine physicians scored patient image quality, and the team measured signal-to-noise and structural similarity against the full-count images.

Half counts held up. Denoised half-count images kept SUVmax values in the phantoms and recovered lesion SUVmax in patients, with noise levels equivalent to full-time images. Readers rated them qualitatively comparable to the full-count scans.

One-third counts went too far. Denoised images at that level were slightly noisier than the originals and risked losing information on lesion detectability. Based on those phantom results, the team reconstructed only 30 patients at one-third counts.

The authors conclude that SubtlePET™ suits half-time or half-dose acquisitions at the European standard of 3 MBq/kg without a loss in diagnostic confidence. The study is useful as much for the line it draws as for the result: a 50% cut held up, and a 66% cut started to cost detail.