SubtlePET™ kept 86.5% of FDG PET lesion texture features concordant with the original, compared with 44.2% for a standard Gaussian filter.
Radiomics models predict outcomes from texture features buried in PET images. For those models, any denoising tool raises a fair question: does cleaning up the image change the data they rely on?
This retrospective study, published in Frontiers in Oncology, measured how SubtlePET™ affected FDG PET texture on a Philips Vereos digital PET/CT, compared with the standard EARL1 Gaussian filter. Researchers in Caen, France, analyzed 113 patients and 101 FDG-avid lesions, plus liver and lung as reference organs. They extracted 104 texture features from the original, AI-denoised, and EARL1-filtered images.
In lesions, 86.5% of features stayed concordant with the original after AI denoising (90 of 104), compared with 44.2% after EARL1 filtering (46 of 104). SUVpeak held steady under both methods. The difference came down to how each method works. AI denoising changed low-intensity regions more than high-intensity ones, while the Gaussian filter changed both the same way.
The study is clear about what did change. Concordance in liver and lung was low for both methods. And in lesions, 79% of features had absolute values that differed significantly from the original, even when they stayed correlated.
The authors frame the result with that in mind. A radiomics model trained on original images could keep the same predictive features after AI denoising, but its thresholds would likely need adjusting. They conclude that AI denoising preserves most lesion texture because it adapts to tissue type, where a fixed Gaussian filter doesn’t.