Quantitative Standardized Uptake Value Evaluation of 4x Faster PET Scans Enhanced using Deep Learning

In 15 patients scanned twice, SubtlePET™-enhanced PET acquired 4x faster matched standard scans on SUVmax and SUVmean, with concordance of 0.99.

December 5, 2019

For PET to be useful in oncology, image quality isn’t the whole story. SUV values have to hold up too, because clinicians track them over time. This study, led by Akshay Chaudhari with Stanford and Subtle Medical colleagues, asked whether SubtlePET™ could keep SUVs accurate on scans acquired 4x faster.

Fifteen patients referred for clinical whole-body PET/CT had two separate PET scans: one at standard acquisition time and one acquired 4x faster. Unlike studies that simulate faster scans from existing data, this one acquired both. The fast scans were enhanced with SubtlePET™. A nuclear medicine physician drew 63 regions of interest on the standard scans, covering possible lesions and normal regions, and those regions were carried over to the enhanced fast scans for comparison.

The SUVs agreed closely. Concordance and Pearson correlation were both 0.99 for SUVmax and SUVmean. Bland-Altman analysis showed minimal differences, with almost all values inside the 95% limits of agreement. Neither measure differed significantly between the two scans (P = 0.68 for SUVmax, P = 0.77 for SUVmean).

The sample was small and one physician drew the regions. Still, it’s direct evidence on the question quantitative PET users ask first: do the numbers change?