Low-count whole-body PET with deep learning in a multicenter and externally validated study

Across 3 hospitals and 50 patients, deep learning made 25%-count whole-body PET noninferior to standard scans for image quality and diagnostic confidence.

February 23, 2021

Cutting PET scan time or radiotracer dose usually costs image quality. Deep learning can help, but most models have only been tested at the sites and on the scanners they were trained on. This study, published in npj Digital Medicine, tested a deep learning enhancement algorithm where it matters: across institutions, scanner types, and patients it had never seen.

Fifty oncology patients at 3 academic hospitals had standard whole-body PET, and the team created fourfold reduced-count versions using only 25% of the counts. Those low-count scans were then enhanced with deep learning. Three board-certified nuclear medicine physicians from different institutions read the standard and enhanced scans blinded.

The enhanced scans were noninferior to standard scans for diagnostic image quality and overall diagnostic confidence, regardless of scanner. Lesion detection on enhanced scans reached a patient-level sensitivity of 0.94 and specificity of 0.98, using standard scans as the reference. SUV values in lesions and reference regions were comparable, with concordance of 0.94 or higher.

The most telling number is about variation. Agreement between enhanced and standard scans (kappa 0.85) was close to how consistently one reader agreed with themselves (0.88), and higher than agreement between different readers (0.72 at most). In other words, swapping in the enhanced scan changed the read less than swapping the reader did.

The authors conclude that deep learning can restore diagnostic image quality and SUV accuracy on fourfold reduced-count PET, and that it generalized to new institutions and scanners. They suggest it could support either a lower dose or a shorter exam.