In 27 patients, sagittal T2 lumbar spine MRI enhanced with SubtleMR™ matched standard image quality at 0:49 instead of 2:12.
Sagittal T2 is a core sequence in lumbar spine MRI. This pilot study, led by Lawrence Tanenbaum at RadNet, tested how far deep learning could shorten it before image quality slipped.
Twenty-seven consecutive patients had standard lumbar spine MRI on one of 3 clinical 1.5T scanners. Each also got an accelerated sagittal T2 processed with SubtleMR™. The standard sagittal T2 averaged 2:12, and the accelerated version averaged 0:49, a 2.7x acceleration. Two board-certified neuroradiologists rated all 54 series in random order for perceived SNR, anatomy and pathology conspicuity, motion artifacts, and overall image quality on a 5-point scale, where 3 meant diagnostic and 5 meant excellent.
Both sets scored high. Reader 1 rated standard scans 4.9 to 5.0 across the four criteria and enhanced scans 4.8 to 4.9. Reader 2’s scores ran from 4.4 to 4.8 for standard scans and 4.2 to 4.7 for enhanced. The enhanced scans scored slightly lower on most measures, but no difference was statistically significant for either reader on any criterion.
The authors conclude that deep learning can match routine lumbar spine image quality at roughly a third of the scan time.