Accelerating Whole-Body PET Acquisitions Using Deep Learning: External validation on foreign country data

On whole-body PET data from Brazil, SubtlePET™ improved image quality metrics on scans up to 4x faster than standard of care.

June 12, 2019

A deep learning model is only as useful as its performance on patients it wasn’t trained on. This pilot study, from Brazilian diagnostics company DASA and Subtle Medical, tested SubtlePET™ on whole-body PET data from a different country and a different clinical setting.

Seven patients referred for FDG PET/CT on a GE Discovery 710 had a standard 2-minute-per-bed scan acquired in list mode. From that data, the team created 2x, 3x, and 4x accelerated versions using only the first 60, 40, and 30 seconds of counts. They enhanced each accelerated scan with SubtlePET™ and compared all of them with the standard scan using NRMSE, PSNR, and SSIM, excluding high-uptake regions like the brain and bladder.

At every acceleration level, the enhanced scans scored significantly better than the non-enhanced ones on all three metrics. The authors also report that enhanced images at 2x through 4x showed perceptual image quality and lesion conspicuity similar to the standard scans.

The sample was small and the faster scans were simulated from the original data, not acquired separately. The authors see the approach as a way to improve patient comfort and PET throughput, particularly in resource-constrained countries.